<thead id="lpzvp"></thead>

    <thead id="lpzvp"></thead>
    <thead id="lpzvp"></thead>

          top
          logo


          Home 派藍尼真空計

          Model Number

          Specifications

          Power/Output

          Features

          PSG 500-S/502-S

          PSG 510-S/512-S

          psg500.jpg

          Pirani Standard Gauge

          *Range: 5X10-4~1000

          *Temp.:

          80/250(Connection)

          5~60(Operating)

          -20~65(Storage)

          *Vacuum Connection:

          DN 16 ISO-KF, 7/16-20 UNF DN16 CF-R, 4 VCR

          1/8” NPT, 1/2” Tube.

          8VCR, DN 16 ISO-KF
            (Long)

          DN16 CF-R Long tube

          *Filament:

          Tungsten & Nickel

          *Supply: 14~30 VDC.

          *Output Signal:

          0~10.3V (Voltage range)

          1.9~10V(Measurement)

          0~0.5V(Error signal)

          10 k0hm (Minimum load)

          *PSG500 without setpoint and the others with two setpoints.

          *Electrical consumption: FCC68/RJ45, 8 poles, male.

          *Cable length: 100m

          *Protection: IP40

          *Easy push button ATM and HV adjustment.

          *Space saving rugged design.

          *Ceramic feedthrough for extremly corrosive applications (PSG510 & PSG512)

          *Stainless steel measuring cell with metal-sealed feedthrough.

          *Logarithmic signal output for easy integration.

          *Nickel filament option for corrosive applications.

          *Aluminium housing.

          Model Number

          Specifications

          Power/Output

          Features

          PVG 500

          PVG 500S/502S

          pvg500s.jpg

          Pirani Standard Gauge

          *Range: 5X10-4~1000

          *Temp.:

          80(Connection)

          5~60(Operating)

          -20~65(Storage)

          *Vacuum Connection:

          KF16

          *Filament:

          Tungsten & Nickel

          *Protection: IP40

          *Cable length: 100m

          *Supply: 14~30 VDC.

          *Output Signal:

          0~10.3V (Voltage range)

          1.9~10V(Measurement)

          0~0.5V(Error signal)

          10 k0hm (Minimum load)

          *PVG500 without setpoint and the others with two setpoints.

          *Electrical consumption: FCC68/RJ45, 8 poles, male.

          *Compact, rugged aluminum housing that mounts in any orientation with a logarithmic signal output.

          *orientation with a logarithmic signal output.

          *Easy push button ATM and High Vacuum (HV) adjustment.

          *Nickel filament option.

          *Solution for corrosive applications

          *Utilize pressure readings to perform critical operations.

          Model Number

          Specifications

          Power/Output

          Features

          PSG 050/051

          psg050.jpg

          Pirani Standard Gauge

          *Range:

          5x10-4~1000 mbar

          *Temp.:

          0~40(Operating)

          80(Storage)

          110(Filament)

           

          *Operating units with vacuum gauge controller VGC012 / VGC023

          *Filament: Tungsten, Platinium.

          *Vacuum connection:

          DN 16 ISO-KF, 1/8” NPT.

          *Tungsten or platinium filament.

          *Constant filament temperature.

          *PSG050: Aluminium sensing cell with tungsten filament. Excellent temperature compensation.

          *PSG050 1/8” NPT: Stainless steel sensing cell with tungsten filament

          *PSG051: Stainless steel sensor with

          platinium filament and ceramics feedthrough. Well suited for corrosive processes and high levels of water vapour.

          Model Number

          Specifications

          Power/Output

          Features

          PSG 100-S/101-S

          psg100.jpg

          Pirani Standard Gauge.

          *Range:

          5x10-4~1000 mbar

          *Temp.:

          80(Bakeout)

          10~50(Operating)

          -20~70(Storage)

          *Filament: Tungsten,Platinum.

          *Supply: 14.5~36 VDC.

          *Output Signal:0~10.6 V

          0.66~10V (Range)

          1.33.V/Decade(Relation V/Pressure)

          *iNTER

          *Cable length: 100m

          *Flange: Tube AD 1/2”, DN16 ISO-KF, DN 16 CF-R, 8VCR.

          *Absolute pressure measurement, independent of gas type.

          *Excellent long-term and temperature stability.

          *Minimal zero drift.

          *Corrosion-resistant alumina construction.

           

          Model Number

          Specifications

          Power/Output

          Features

          PCG 400/410

          PCG 400-S/410-S

          pcg400.jpg

          Pirani Capacitance Diaphragm Gauge

          *Range: 5x10-4~1500 mbar

          *Temp.:

          80(Bakeout)

          10~50(Operating)

          -20~65(Storage)

          <160(Filament)

          *Connector: FCC 68, female, 8 pin(shielded).

          *PCG400-S with 2 setpoints FCC 68 and PCG410-S with 1 setpoint 9 D-Sub.

          *Supply: 15~30 VDC. 2.5W

          *Output Signal:

          0~10.3 V(400/400S)

           0~9 V (410/410S)

          *Sensor cable length: 100m

          *Flange: DN 16 CF-F, 16 ISO-KF,

          *Fitting: 8 VCR(Female), 1/8” NPT, 4 VCR(Female), 4 VCR 900 Female.

          *Protection: IP40

           

          *Gas-type-independent above 100 mbar (75 Torr) - enables safe venting with any gas mixture

          *Higher accuracy and reproducibility at atmosphere.

          *Independence of mounting position.

          *Faster atmosphere detection.

          *Compact design.

          *One log/linear output signal over the entire range.

          Model Number

          Specifications

          Power/Output

          Features

          PCG 550/552/554

          pcg550.jpg

          Pirani Capacitance Diaphragm Gauge

          *Range: 5x10-5~1500 mbar

          *Temp.:

          80(Bakeout)

          10~50(Operating)

          -20~65(Storage)

          250(Long tube)

          *Flange: DN 16 ISO KF or Long tube, DN16 CF-F (or 16 CF-R Long tube), DN25 ISO KF,4VCR

          *Fitting:8 VCR, 1/8” NPT, 7/16-20 UNF, 29X29 mm,4 VCR 90O.

          *Supply: 15~30 VDC.

          *Output Signal:0~10.23V,

          0~8.68V,0~5.659,0~9.05.

          *Interface: RS232C(Standard),DeviceNet,Profibus DP, RS485C.

          *Electrical connection:

          FCC, 8 pin, D-Sub 9 pin/ 15 pin HD male, Hirschmann, 6 pin.

          *Cable length:100m,

          30m(RS232C)

          *Filament: Tungsten, Nickel, Ceramic coated.

          *Protection: IP40.

          *Gas-type independent above 10 mbar.

          *High accuracy and reproducibility at atmosphere.

          *Versatile of mounting orientation.

          *Available with Tungsten (PCG550) or Nickel (PCG552) filament or with a fully ceramic coated (PCG554) sensor unit for highly corrosive applications.

          *Easy to exchange plug & play sensor

          element with on-board calibration 
             Data.

          *Eligible output signal for easy integration.

          *Optional atmospheric switch, display and digital interfaces.

          Model Number

          Specifications

          Power/Output

          Features

          PCG 750 /752

          pcg750522.jpg

          Pirani Capacitance Diaphragm Gauge

          *Range:

          5x10-5~1500 mbar

          *Temp.:

          80(Bakeout)

          10~50(Operating)

          -20~65(Storage)

          *Flange: KF16

          *Supply: 15~30 VDC. 2 W

          *Output Signal:0~10.23 V

          0.61~10.23V (Measurement)

          1.286 V/decade (Voltage vs. pressure)

          >10 k0hm (Load impedance)

          *Interface: RS232C.

          *Sensor cable length: 500m

          *Flange: KF16

          *Power connection FCC, 8 pin.

          *High accuracy and reproducibility at atmosphere.

          *Fast atmospheric detection.

          *Gas-type independent above 10
             mbar.

          *Nickel filament option.

          * Compact, rugged housing that mounts in any orientation with a logarithmic signal output.

          pcg75052.jpg

           

          bottom

          Copyright © 2007 Best Champion Technology Co., Ltd. , All Rights Reserved. Tel: 03-561-0288 Fax: 03-561-0878 E-mail: sales@bcv.com.tw